K-Lite8 Series Electrolyte Analyzer (Auto Sampler is optional)
K-Lite8A: K+, Na+, Cl-, TCO2;
K-Lite8B: K+, Na+, Cl-, Ca++, TCO2, pH;
K-Lite8D: K+, Na+, Cl-, Ca++, Li+, TCO2, pH;
K-Lite8E: K+, Na+, Cl-,
K-Lite8G: K+, Na+, Cl-, Ca++, pH;
K-Lite8F: K+, Na+, Li+;
K-Lite8H: K+, Na+, Cl-, Ca++, Li+, pH;
 Sample Type

Plasma, Serum, Whole Blood, Urine

Parameter Configuration

K-Lite8A:  K+, Na+, Cl-, TCO2;

K-Lite8B:  K+, Na+, Cl-, Ca++, TCO2, pH;

K-Lite8D:  K+, Na+, Cl-, Ca++,  Li+, TCO2, pH;

K-Lite8E:  K+, Na+, Cl-,

K-Lite8G:  K+, Na+, Cl-, Ca++, pH;

K-Lite8F:  K+, Na+, Li+;

K-Lite8H:  K+, Na+, Cl-, Ca++,  Li+, pH;

Auto Sampler is optional

Test Range and Accuracy

Item

Range (mmol/L)

SD(mmol/L)

CV%

K

0.5-15.5

±0.2

≤1.0

Na

20.0-200.0

±2..0

≤1.0

Cl

20.0-200.0

±2.5

≤1.0

Li

0.1~3.0

±0.1

≤1.5

Ca

0.1-5.0

±0.1

≤1.5

pH

4.0-9.0

±0.1

≤1.0

TCO2

6.0-50.0

±2.0

≤3.0

Working Environment

Ambient Temperature

5℃-40℃

Relative Humidity

≦85%

Sun Radiation

avoid direct sunlight

Power Supply

100-240V~50/60Hz

Main Characteristics:

Ø  7” colorful touch screen and interactive menu

Ø  Real time display of testing status and reviewing test results

Ø  Random parameters combination of K, Na, Cl, Ca, pH, Li, TCO2

Ø  Standard LAN interface with HL7 protocol and dual-direction LIS transmission

Ø  Powerful QC program and QC management

Ø  Optional auto sampler applicable for cups and sample tubes

Ø  Liquid level detect and probe collision protection

Ø  Barcode managed closed reagent system

Ø  Sample volume: 65ul—150ul