You are here:Home > Products
Service hotline
Tel:+86-755-86330432
Fax:+86-755-86330882
Email:sales@cornley.com
K-Lite6 Series Electrolyte Analyzer (Auto Sampler is optional)
K-Lite6A: K+, Na+, Cl-, TCO2;
K-Lite6B: K+, Na+, Cl-, Ca++, TCO2, pH;
K-Lite6D: K+, Na+, Cl-, Ca++, Li+, TCO2, pH;
K-Lite6E: K+, Na+, Cl-,
K-Lite6G: K+, Na+, Cl-, Ca++, pH;
K-Lite6F: K+, Na+, Li+;
K-Lite6H: K+, Na+, Cl-, Ca++, Li+, pH;
 

Sample Type

Plasma, Serum, Whole Blood, Urine

Parameter Configuration

K-Lite6A:  K+, Na+, Cl-, TCO2;

K-Lite6B:  K+, Na+, Cl-, Ca++, TCO2, pH;

K-Lite6D:  K+, Na+, Cl-, Ca++,  Li+, TCO2, pH;

K-Lite6E:  K+, Na+, Cl-,

K-Lite6G:  K+, Na+, Cl-, Ca++, pH;

K-Lite6F:  K+, Na+, Li+;

K-Lite6H:  K+, Na+, Cl-, Ca++,  Li+, pH;

Auto Sampler is optional

Test Range and Accuracy

Item

Range (mmol/L)

SD(mmol/L)

CV%

K

0.5-15.5

±0.2

≤1.0

Na

20.0-200.0

±2..0

≤1.0

Cl

20.0-200.0

±2.5

≤1.0

Li

0.1~3.0

±0.1

≤1.5

Ca

0.1-5.0

±0.1

≤1.5

pH

4.0-9.0

±0.1

≤1.0

TCO2

6.0-50.0

±2.0

≤3.0

Working Environment

Ambient Temperature

5℃-40℃

Relative Humidity

≦85%

Sun Radiation

avoid direct sunlight

Power Supply

100-240V~50/60Hz

Main Characteristics:

Ø  Maintenance - free electrodes with long shelf life

Ø  Closed reagent system to avoid biological contamination

Ø  Unique consumable identification system

Ø  Optional auto sampler applicable for cups and sample tubes

Ø  Liquid detection for auto sampler

Ø  Probe collision protection

Ø  Optional sample bar code reader

Ø  Rapid test cycle: 90 seconds

Ø  Sample volume:150µL